![Rohde & Schwarz FSW13 Скачать руководство пользователя страница 760](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847760.webp)
Remote Commands
R&S
®
FSW
760
User Manual 1173.9411.02 ─ 19
Parameters:
<Bandwidth>
Bandwidth in Hz
*RST:
10.0 MHz
Example:
//Select the band category 1
ESP2:MSR:BCAT BC1
//Set the base station class to medium range
ESP2:MSR:CLAS MED
//Set the maximum output power to 10 dBm.
ESP2:MSR:MPOW 10
//Set the frequency range of the base station to > 3 GHz
ESP2:MSR:BAND:HIGH
//Set the base station RF bandwidth to 20 MHz
ESP2:MSR:RFBW 20MHZ
//Calculate limits for MSR SEM
ESP2:MSR:APPL
Manual operation:
See
11.5.6.8
Configuring the List Evaluation
The following commands configure the list evaluation.
Useful commands for SEM measurements described elsewhere
●
Remote commands exclusive to SEM measurements
CALCulate<n>:ESPectrum:PSEarch|PEAKsearch:AUTO
...................................................760
CALCulate<n>:ESPectrum:PSEarch|PEAKsearch[:IMMediate]
........................................... 760
CALCulate<n>:ESPectrum:PSEarch|PEAKsearch:MARGin
................................................761
CALCulate<n>:ESPectrum:PSEarch|PEAKsearch:PSHow
................................................. 761
CALCulate<n>:ESPectrum:PSEarch|PEAKsearch:AUTO
<State>
This command turns the list evaluation on and off.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
CALC:ESP:PSE:AUTO OFF
Deactivates the list evaluation.
Manual operation:
See
"List Evaluation (Result Summary) State"
See
CALCulate<n>:ESPectrum:PSEarch|PEAKsearch[:IMMediate]
This command initiates a list evaluation.
Configuring and Performing Measurements