General Instrument Setup
R&S
®
FSW
576
User Manual 1173.9411.02 ─ 19
Save
Saves the currently edited transducer line under the name defined in the "Name" field.
Remote command:
9.3.3 Alignment Settings
Both the instrument and the touchscreen can be aligned when necessary (see
ter 9.1, "Basics on Alignment"
The alignment settings are defined in the "Alignment" dialog box which is displayed
when you press the SETUP key and then select "Alignment".
Self-alignment results in secure user mode
Be sure to store self-alignment results before
ter 3.1.7, "Protecting Data Using the Secure User Mode"
If the currently stored self-alignment results are not suitable, you must perform a self-
alignment each time you switch on the R&S
FSW in secure user mode, as the results
are not stored permanently in this case.
.............................................................................................576
.........................................................................................577
Starting the Touchscreen Alignment
........................................................................... 577
Starting a Self-alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details see
chapter 9.1, "Basics on Alignment"
Note:
General Instrument Settings