Remote Commands
R&S
®
FSW
695
User Manual 1173.9411.02 ─ 19
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see
chapter 10.1.6, "Command Sequence and Synchro-
For details on synchronization see the "Remote Basics" chapter in the R&S
FSW User
Manual.
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Sets the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
Manual operation:
See
See
See
[SENSe:]SWEep:COUNt:CURRent?
This query returns the current number of started sweeps or measurements. This com-
mand is only available if a sweep count value is defined and the instrument is in single
sweep mode.
Example:
SWE:COUNt 64
Sets sweep count to 64
INIT:CONT OFF
Switches to single sweep mode
INIT
Starts a sweep (without waiting for the sweep end!)
SWE:COUN:CURR?
Queries the number of started sweeps
Usage:
Query only
11.5.2 Configuring Power Measurements
The following commands work for several power measurements.
CALCulate<n>:MARKer<m>:FUNCtion:POWer:MODE
...................................................... 696
CALCulate:MARKer:FUNCtion:POWer<sb>:RESult?
.........................................................696
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect
..................................................... 698
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
.................................................... 698
Configuring and Performing Measurements