Remote Commands
R&S
®
FSW
866
User Manual 1173.9411.02 ─ 19
INPut:FILTer:YIG[:STATe]
<State>
This command turns the YIG-preselector on and off.
Note the special conditions and restrictions for the YIG filter described in
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1 (0 for I/Q Analyzer, GSM, VSA and MC Group
Delay measurements)
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation:
See
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input.
75
Ω
should be selected if the 50
Ω
input impedance is transformed to a higher impe-
dance using a matching pad of the RAZ type (= 25
Ω
in series to the input impedance
of the instrument). The power loss correction value in this case is 1.76 dB = 10 log
(75
Ω
/50
Ω
).
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Example:
INP:IMP 75
Usage:
SCPI confirmed
Manual operation:
See
See
INPut:SELect
<Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S
FSW. If no additional options are installed, only
RF input is supported.
Setting Basic Measurement Parameters