Remote Commands
R&S
®
FSW
847
User Manual 1173.9411.02 ─ 19
INPut:EATT
<Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (
INP:EATT:AUTO OFF
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
Parameters:
<Attenuation>
attenuation in dB
Range:
see data sheet
Increment: 1 dB
*RST:
0 dB (OFF)
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation:
See
"Using Electronic Attenuation (Option B25)"
INPut:EATT:AUTO
<State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:EATT:AUTO OFF
Manual operation:
See
"Using Electronic Attenuation (Option B25)"
INPut:EATT:STATe
<State>
This command turns the electronic attenuator on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:EATT:STAT ON
Switches the electronic attenuator into the signal path.
Manual operation:
See
"Using Electronic Attenuation (Option B25)"
11.7.3.3
Configuring a Preamplifier
Setting Basic Measurement Parameters