Measurements
R&S
®
FSW
190
User Manual 1173.9411.02 ─ 19
5.5.2 OBW Results
As a result of the OBW measurement the occupied bandwidth ("Occ BW") is indicated
in the marker results. Furthermore, the marker at the center frequency and the tempo-
rary markers are indicated.
The measurement is performed on the trace with marker 1. In order to evaluate
another trace, marker 1 must be placed on another trace (see
The OBW calculation is repeated if the
are changed, without performing
a new sweep. Thus, the OBW for a multicarrier signal can be determined using only
one sweep.
Remote command:
The determined occupied bandwidth can also be queried using the remote command
CALC:MARK:FUNC:POW:RES? OBW
or
CALC:MARK:FUNC:POW:RES? AOBW
. While
the
OBW
parameter returns only the occupied bandwidth, the
AOBW
parameter also
returns the position and level of the temporary markers T1 and T2 used to calculate the
occupied bandwidth.
CALC:MARK:FUNC:POW:SEL OBW
, see
CALC:MARK:FUNC:POW:RES? OBW
, see
Occupied Bandwidth Measurement (OBW)