Data Management
R&S
®
FSW
549
User Manual 1173.9411.02 ─ 19
Save File
Saves the settings file with the defined file name.
Note:
Secure user mode. In secure user mode, settings that are to be stored on the
instrument are stored to volatile memory, which is restricted to 256
MB. Thus, a "Mem-
ory full" error may occur although the hard disk indicates that storage space is still
available.
For details see
chapter 3.1.7, "Protecting Data Using the Secure User Mode"
Remote command:
Recall in New Channel / Recall in Current Channel
Restores the instrument settings as saved in the selected settings file. If the settings
file contains settings for a specific channel only, select "Recall in New Channel" to acti-
vate a new channel with the stored settings, or "Recall in Current Channel" to replace
the current channel settings.
Note:
After you use the "Recall" function, the history of previous actions is deleted, i.e.
any actions performed previously cannot be undone or redone using the UNDO/REDO
keys.
Note:
If a measurement channel with the same name as the channel to be restored (in
a new channel) is already active, the channel name for the new channel is extended by
a consecutive number:
In remote commands, you must append this number to the channel name, as well.
Remote command:
Cancel
Closes the dialog box without saving the settings.
8.3.2.4
Startup Recall Settings
The "Startup Recall" softkey opens the "Startup Recall" tab of the "Recall" dialog box.
Storing and Recalling Instrument Settings and Measurement Data