![Rohde & Schwarz FSW13 Скачать руководство пользователя страница 705](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847705.webp)
Remote Commands
R&S
®
FSW
705
User Manual 1173.9411.02 ─ 19
Parameters:
<Number>
Range:
1 to 18
*RST:
1
Manual operation:
See
"Number of Channels (Tx, ADJ)"
11.5.3.3
Defining Weighting Filters
The following commands define weighting filters for ACLR measurements.
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
....................................................... 705
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
...............................................................705
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
................................................705
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>
................................................. 706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
......................................................706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALL
...............................................................706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
.............................................. 706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
................................................706
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
<Alpha>
This command defines the roll-off factor for the adjacent channel weighting filter.
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
<Value>
This command defines the alpha value for the weighting filter for all channels.
Parameters:
<Value>
<numeric value>
*RST:
0.22
Example:
POW:ACH:FILT:ALPH:ALL 0.35
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
<Alpha>
This command defines the roll-off factor for the alternate channel weighting filter.
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
See