[SENSe:]SWEep:EGATe:TYPE
<Type>
This command selects the way gated measurements are triggered.
Parameters:
<Type>
LEVel
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open
until the signal disappears.
Note:
If you perform gated measurements in combination with
the IF Power trigger, the R&S
FSW ignores the holding time for
frequency sweep, FFT sweep, zero span and I/Q mode meas-
urements.
EDGE
The trigger event for the gate to open is the detection of the sig-
nal edge.
After the gate signal has been detected, the gate remains open
until the gate delay is over.
*RST:
EDGE
Example:
SWE:EGAT:TYPE EDGE
Manual operation:
See
11.7.4.3
Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
TRIGGER INPUT/OUTPUT connectors. The tasks for manual operation are described
in
OUTPut:TRIGger<port>:DIRection
...................................................................................858
........................................................................................ 859
...................................................................................... 859
OUTPut:TRIGger<port>:PULSe:IMMediate
.......................................................................860
OUTPut:TRIGger<port>:PULSe:LENGth
.......................................................................... 860
OUTPut:TRIGger<port>:DIRection
<Direction>
This command selects the trigger direction.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)