Remote Commands
R&S
®
FSW
859
User Manual 1173.9411.02 ─ 19
Parameters:
<Direction>
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Manual operation:
See
OUTPut:TRIGger<port>:LEVel
<Level>
This command defines the level of the signal generated at the trigger output.
This command works only if you have selected a user defined output with
.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Level>
HIGH
TTL signal.
LOW
0 V
*RST:
LOW
Manual operation:
See
See
OUTPut:TRIGger<port>:OTYPe
<OutputType>
This command selects the type of signal generated at the trigger output.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Setting Basic Measurement Parameters