Remote Commands
R&S
®
FSW
815
User Manual 1173.9411.02 ─ 19
<TriggerSource>
EXTernal | EXT2 | EXT3 | IMMediate | IFPower | RFPower |
VIDeo
Selects a trigger source.
For more information see
Configuring Triggered and Gated
.
<TriggerSlope>
NEGative | POSitive
Selects the trigger slop.
<TriggerOffset>
Defines the trigger delay.
Range:
negative measurement time to 30
*RST:
0
Default unit: s
<GateLength>
Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off.
To perform gated measurements, the trigger source must be dif-
ferent from
IMMediate
.
Range:
31.25 ns to 30 s
*RST:
0 s
[SENSe:]LIST:POWer:STATe
<State>
This command turns the List Evaluation off.
Parameters:
<State>
OFF
*RST:
OFF
11.5.14.2
Example: Performing List Evaluation
The following example shows a list evaluation with the following configuration.
No
Freq
[MHz]
Ref
Level
[dBm]
RF
Attenu-
ation
[dB]
El
Attenu-
ation
[dB]
Filter
RBW
VBW
Meas
Time
Trigger
Level
1
935.2
0
10
---
Normal
1 MHz
3 MHz
440 µs
0
2
935.4
0
10
10
Channel 30 kHz
100 kHz
440 µs
0
3
935.6
0
10
20
Channel 30 kHz
100 kHz
440 µs
0
-----Measurement with synchronization via service request-----
*ESE 1
*SRE 32
// Configures the status reporting system to produce a service request.
LIST:POW:SET ON,ON,OFF,EXT,POS,10us,434us
//Turns on the list evaluation, configures the global list evaluation settings and
//evaluates the peak and RMS power.
Configuring and Performing Measurements