Parameters:
<Position>
0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST:
100 PCT = frequency display; 50 PCT = time dis-
play
Example:
DISP:TRAC:Y:RPOS 50PCT
Usage:
SCPI confirmed
Manual operation:
See
See
DISPlay[:WINDow<n>]:TRACe:Y:SPACing
<ScalingType>
This command selects the scaling of the y-axis.
Parameters:
<ScalingType>
LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST:
LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Usage:
SCPI confirmed
Manual operation:
See
11.7.4 Configuring Triggered and Gated Measurements
The commands required to configure a triggered or gated measurement in a remote
environment are described here. The tasks for manual operation are described in
chapter 6.6, "Trigger and Gate Configuration"
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions
...................................................................851
Configuring Gated Measurements
........................................................................ 856
Configuring the Trigger Output
..............................................................................858
Setting Basic Measurement Parameters