
Remote Commands
R&S
®
FSW
694
User Manual 1173.9411.02 ─ 19
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sweeps) and waits for the
end.
Result: Averaging is performed over 40 sweeps.
Manual operation:
See
INITiate:CONTinuous
<State>
This command controls the sweep mode.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see
chapter 10.1.6, "Command Sequence and Synchro-
If the sweep mode is changed for a measurement channel while the Sequencer is
active (see
on page 688) the mode is only con-
sidered the next time the measurement in that channel is activated by the Sequencer.
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
Example:
INIT:CONT OFF
Switches the sweep mode to single sweep .
INIT:CONT ON
Switches the sweep mode to continuous sweep .
Manual operation:
See
See
See
INITiate[:IMMediate]
This command starts a (single) new measurement.
Configuring and Performing Measurements