General Instrument Setup
R&S
®
FSW
603
User Manual 1173.9411.02 ─ 19
9.6.3 How to Align the Touchscreen
To align the touchscreen
1. Press the SETUP key.
2. Select the "Alignment" softkey.
3. Select "Touchscreen Alignment".
A blinking cross appears in the lower left corner of the screen.
4. Touch and hold the blinking cross until it stops blinking.
Repeat this action for the crosses in the other corners.
9.6.4 How to Install an R&S
FSW Option
Additional options for the R&S
FSW can be enabled using a license key. To obtain the
license key, consult your sales representative. You need the device ID and serial num-
ber of your instrument to get a license key (see
chapter 9.3.4, "System Configuration
on page 577). No additional installation is required.
9.6.5 How to Update the Instrument Firmware
1. Download the update package from the Rohde&Schwarz website and store it on a
memory stick, on the instrument, or on a server network drive that can be accessed
by the instrument.
2.
NOTICE!
Stop measurement. The firmware update must not be performed during a
running measurement.
If a measurement is running, stop it by pressing the highlighted RUN CONT or
RUN SINGLE key.
3. Press the SETUP key.
4. Select the "Firmware Update" tab.
5. In the file selection dialog box select the
FSWSetup*.exe
file.
6. Select "Install" to start the update.
7. After the firmware update, the R&S
FSW reboots automatically.
8. Depending on the previous firmware version, a reconfiguration of the hardware
might be required during the first startup of the firmware. The reconfiguration starts
automatically, and a message box informs you about the process. When the recon-
figuration has finished, the instrument again reboots automatically.
How to Configure the Basic Instrument Settings