Remote Commands
R&S
®
FSW
711
User Manual 1173.9411.02 ─ 19
Parameters:
<LowerLimit>,
<UpperLimit>
The first value defines the limit of the lower alternate channel,
the second value the limit of the upper alternate channel.
Range:
0 dB to 100 dB
*RST:
0 DB
Default unit: dB
Manual operation:
See
See
CALCulate:LIMit:ACPower:ALTernate<ch>:RESult?
This command queries the state of the limit check for the adjacent or alternate chan-
nels in an ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Return values:
<LowerChan>,
<UpperChan>
text value
The command returns two results. The first is the result for the
lower, the second for the upper adjacent or alternate channel.
PASSED
Limit check has passed.
FAIL
Limit check has failed.
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage:
Query only
CALCulate:LIMit:ACPower:ALTernate<ch>[:RELative]:STATe
<State>
This command turns the relative limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
See
CALCulate:LIMit:ACPower[:STATe]
<State>
This command turns the limit check for ACLR measurements on and off.
Configuring and Performing Measurements