3.1.1.6
Performing a Self-Alignment and a Selftest
During instrument start, the installed hardware is checked against the current firmware
version to ensure the hardware is supported. If not, an error message is displayed
("WRONG_FW") and you are asked to update the firmware. Until the firmware version
is updated, self-alignment fails.
(For details refer to the R&S
FSW User Manual).
Furthermore, it may be necessary to align the data to a reference source by performing
a self-alignment when strong temperature changes occur.
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
A message in the status bar ("Instrument warming up...") indicates that the operating
temperature has not yet been reached.
Performing a self-alignment
1. Press the SETUP key.
2. Press the "Alignment" softkey.
3. Select the "Start Self-Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
●
Press the SETUP key.
●
Press the "Alignment" softkey.
Performing a selftest
The selftest does not need to be repeated every time the instrument is switched on. It
is only necessary when instrument malfunction is suspected.
1. Press the SETUP key.
2. Press the "Service" softkey.
3. Switch to the "Selftest" tab in the "Service" dialog box.
4. Select the "Start Selftest" button.
Once the instrument modules have been checked successfully, a message is dis-
played.
Preparing for Use