Remote Commands
R&S
®
FSW
1019
User Manual 1173.9411.02 ─ 19
[SENSe:]ROSCillator:LBWidth
<Bandwidth>
Defines the loop bandwidth, that is, the speed of internal synchronization with the refer-
ence frequency. The setting requires a compromise between performance and increas-
ing phase noise.
For a variable external reference frequency with a narrow tuning range (+/- 0.5 ppm),
the loop bandwidth is fixed to 0.1 Hz and cannot be changed.
Parameters:
<Bandwidth>
0.1 Hz | 1 Hz | 3 Hz | 10 Hz | 30 Hz | 100 Hz | 300 Hz
The possible values depend on the reference source and tuning
range (see
).
Default unit: Hz
Example:
ROSC:LBW 3
Manual operation:
See
[SENSe:]ROSCillator:O100
<State>
If enabled, a 100 MHz reference signal is provided to the REF OUTPUT 100 MHZ con-
nector.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
ROSC:O100 ON
Manual operation:
See
[SENSe:]ROSCillator:O640
<State>
If enabled, a 640 MHz reference signal is provided to the REF OUTPUT 640 MHZ con-
nector.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
ROSC:O640 ON
Manual operation:
See
SOURce:EXTernal:ROSCillator:EXTernal:FREQuency
<Frequency>
This command defines the frequency of the external reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Parameters:
<Frequency>
Range:
1 MHz to 20 MHz
Configuring the R&S
FSW