Measurements
R&S
®
FSW
301
User Manual 1173.9411.02 ─ 19
If a trace is turned off, the assigned markers and marker functions are also deactiva-
ted.
Remote command:
Final Test Detector
Defines the detector to be used for the final EMI test at the marker frequency. This set-
ting is only available if the EMI (R&S
FSW-K54) measurement option is installed.
For details see
chapter 5.13.3.2, "Detectors and Dwell Time"
Note:
The trace detector configured in the trace settings is used for the initial peak
chapter 7.3.2.1, "Trace Settings"
"Off"
No final test is performed.
"Positive peak" Determines the maximum signal level that was detected during the
specified dwell time.
"Average"
Determines the average signal level of the samples that were collec-
ted during the specified dwell time.
"Quasipeak"
Determines the maximum signal level weighted to CISPR
16-1-1 that
was detected during the dwell time.
The quasipeak detector is only available for the CISPR filter, and not
for an RBW of 1
MHz.
"CISPR Aver-
age"
Determines a weighted average signal level according to CISPR
16-1-1.
The average value according to CISPR 16-1-1 is the maximum value
detected while calculating the linear average value during the speci-
fied dwell time.
The CISPR Average detector is only available for the CISPR filter.
"RMS Aver-
age"
A combination of the RMS detector (for pulse repetition frequencies
above a corner frequency) and the Average detector (for pulse repeti-
tion frequencies below the corner frequency).
The average value is determined by lowpass filters of the 2nd order
(simulation of a mechanical pointer instrument).
The RMS Average detector is only available for the CISPR filter.
Remote command:
Select Marker
Opens a dialog box to select and activate or deactivate one or more markers quickly.
Electromagnetic Interference (EMI) Measurement (R&S
FSW-K54)