![Rohde & Schwarz FSW13 Скачать руководство пользователя страница 652](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847652.webp)
Network and Remote Operation
R&S
®
FSW
652
User Manual 1173.9411.02 ─ 19
"New"
R&S
FSW format
Remote command:
I/O Logging
Activates or deactivates the SCPI error log function. All remote control commands
received by the R&S
FSW are recorded in the following log file:
C:\R_S\instr\ScpiLogging\ScpiLog.txt
Logging the commands may be extremely useful for debug purposes, e.g. in order to
find misspelled keywords in control programs.
Remote command:
10.4.3 Compatibility Settings
The R&S
FSW can emulate the GPIB interface of other signal and spectrum analyzers,
e.g. in order to use existing control applications.
Compatibility with former R&S signal and spectrum analyzers
As a rule, the R&S
FSW supports most commands from previous R&S signal and
spectrum analyzers such as the FSQ, FSP, FSU, or FSV. However, the default values,
in particular the number of sweep points or particular bandwidths, may vary. Therefore,
the R&S
FSW can emulate these other devices, including their default values, in order
to repeat previous measurements or support existing control applications as in legacy
systems.
The required settings are configured in the "Compatibility" tab of the "Network
+Remote" dialog box.
Network and Remote Control Settings