Remote Commands
R&S
®
FSW
1041
User Manual 1173.9411.02 ─ 19
Example:
SYST:ERR:EXT? FAT,'Spectrum2'
Queries fatal errors in the 'Spectrum2' application. If none have
occurred, the result is:
"
"
.
Usage:
Query only
SYSTem:ERRor:LIST?
This command queries the system messages.
Return values:
<Messages>
String containing all messages in the "System Messages" table.
Example:
SYST:ERR:LIST?
Usage:
Query only
SYSTem:ERRor[:NEXT]?
This command queries the most recent error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
For details on error queues see
chapter 10.1.7, "Status Reporting System"
Usage:
Query only
SYSTem:FORMat:IDENt
<IDNFormat>
This command selects the response format to the
Parameters:
<IDNFormat>
LEGacy
Format is compatible to R&S FSP/FSU/FSQ/FSG family.
NEW | FSL
R&S
FSW format
Format is also compatible to the R&S FSL and R&S FSV family
*RST:
not reset!
Example:
SYST:FORM:IDEN LEG
Adapts the return value of *IDN? to the R&S FSP/FSU/FSQ fam-
ily.
Manual operation:
See
SYSTem:PRESet:COMPatible
<OpMode>
This command defines the operating mode that is activated when you switch on the
R&S
FSW or press the PRESET key.
Configuring the R&S
FSW