![Rohde & Schwarz FSW13 Скачать руководство пользователя страница 798](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847798.webp)
Remote Commands
R&S
®
FSW
798
User Manual 1173.9411.02 ─ 19
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST?
This command queries the position of the harmonics.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Return values:
<Harmonics>
Returns one value for every harmonic.
The first value is the absolute power of the first harmonic. The
unit is variable.
The other values are power levels relative to the first harmonic.
The unit for these is dB.
Usage:
Query only
Manual operation:
See
11.5.10.5
Example: Measuring the Harmonic Distortion
//-------------Configuring the Measurement-------------
*RST
//Resets the instrument.
INIT:CONT OFF
//Turns on single sweep mode.
CALC:MARK:FUNC:HARM ON
//Turns on the harmonic distortion measurement.
CALC:MARK:FUNC:HARM:NHAR 3
//Defines three harmonics to be found.
CALC:MARK:FUNC:HARM:BAND:AUTO OFF
//Turns off automatic bandwidth selection.
CALC:MARK:FUNC:HARM:PRES
//Determines the ideal configuration.
//-------------Performing the Measurement-------------------
INIT;*WAI
//Initiates the measurement and finishes the sweep.
//-------------Retrieving the Results------------------------
CALC:MARK:FUNC:HARM:LIST?
//Queries the position of the harmonics.
Configuring and Performing Measurements