Measurements
R&S
®
FSW
274
User Manual 1173.9411.02 ─ 19
monic Distortion" dialog box, which is displayed as a tab in the "Analysis" dialog box or
when you select the "Harmonic Distortion Config" softkey from the "Harm Dist" menu.
The remote commands required to perform these tasks are described in
ter 11.5.10, "Measuring the Harmonic Distortion"
................................................................................................274
No. of Harmonics
Defines the number of harmonics to be measured. The range is from 1 to 26. Default is
10.
Remote command:
Harmonic Sweep Time
Defines the sweep time for the zero span measurement on each harmonic frequency.
This setting is identical to the normal sweep time for zero span, see also
Remote command:
Harmonic RBW Auto
Enables/disables the automatic adjustment of the resolution bandwidth for Normal
(3dB) (Gaussian) and 5-Pole filter types. The automatic adjustment is carried out
according to:
"RBW
n
= RBW
1
* n"
If RBW
n
is not available, the next higher value is used.
Remote command:
Harmonic Distortion Measurement