
Remote Commands
R&S
®
FSW
840
User Manual 1173.9411.02 ─ 19
Remote commands exclusive to configuring sweeps:
.................................................................................................840
.............................................................................................840
.................................................................................................841
......................................................................................... 842
.........................................................................................843
[SENSe:]BANDwidth|BWIDth[:RESolution]:FFT
.................................................................843
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweeps, the application calculates the moving average over the
average count.
In case of single sweeps, the application stops the measurement and calculates the
average after the average count has been reached.
Parameters:
<SweepCount>
When you set a sweep count of 0 or 1, the R&S
FSW performs
one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Usage:
SCPI confirmed
Manual operation:
See
[SENSe:]SWEep:OPTimize
<Mode>
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called
partial span
. The
partial span cannot be defined directly, but it can be optimized according to measure-
ment requirements.
Setting Basic Measurement Parameters