Measurements
R&S
®
FSW
150
User Manual 1173.9411.02 ─ 19
Setting a Fixed Reference for Channel Power Measurements (Set CP Reference)
Optimized Settings (Adjust Settings)
...........................................................................154
Standard
The main measurement settings can be stored as a standard file. When such a stand-
ard is loaded, the required channel and general measurement settings are automati-
cally set on the R&S
FSW. However, the settings can be changed. Predefined stand-
ards are available for standard measurements, but standard files with user-defined
configurations can also be created.
Note:
If the "Multi-Standard Radio" standard is selected, the "ACLR Setup" dialog box
is replaced by the "MSR ACLR Setup" dialog box (see
If any other predefined standard (or "NONE") is selected, the "ACLR Setup" dialog box
is restored (see
chapter 5.3.4, "Channel Power Configuration"
Note that changes in the configuration are not stored when the dialog boxes are
exchanged.
Predefined Standards ← Standard
Predefined standards contain the main measurement settings for standard measure-
ments. When such a standard is loaded, the required channel settings are automati-
cally set on the R&S
FSW. However, the settings can be changed.
The predefined standards contain the following settings:
●
Channel bandwidths
●
Channel spacings
●
Detector
●
Trace Average setting
●
Resolution Bandwidth (RBW)
●
Weighting Filter
Predefined standards can be selected via the "CP/ACLR Standard" softkey in the "CH
Power" menu or in the "General Settings" tab of the "CP/ACLR Setup" dialog box.
For details on the available standards see
chapter 5.3.9, "Reference: Predefined CP/
Remote command:
User-Defined Standards ← Standard
In addition to the predefined standards you can save your own standards with your
specific measurement settings in an xml file so you can use them again at a later time.
User-defined standards are stored on the instrument in the
C:\R_S\instr\acp_std
directory.
A sample file is provided for an MSR ACLR measurement (MSR_ACLRExample.xml).
It sets up the measurement for the MSR signal generator waveform described in the
file
C:\R_S\instr\user\waveform\MSRA_GSM_WCDMA_LET_GSM.wv
.
Note that ACLR user standards are not supported for Fast ACLR and multicarrier
ACLR measurements.
Channel Power and Adjacent-Channel Power (ACLR) Measurement