[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
<Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor>
1 to 100 PCT
*RST:
10
Example:
FREQ:CENT:STEP:LINK:FACT 20PCT
Manual operation:
See
[SENSe:]FREQuency:OFFSet
<Offset>
This command defines a frequency offset.
If this value is not 0
Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also
Parameters:
<Offset>
Range:
-100 GHz to 100 GHz
*RST:
0 Hz
Example:
FREQ:OFFS 1GHZ
Usage:
SCPI confirmed
Manual operation:
See
[SENSe:]FREQuency:SPAN
<Span>
This command defines the frequency span.
If you set a span of 0
Hz in the Spectrum application, the R&S
FSW starts a measure-
ment in the time domain.
Parameters:
<Span>
The minimum span for measurements in the frequency domain
is 10
Hz. For SEM and spurious emission measurements, the
minimum span is 20
Hz.
Range:
0 Hz to fmax
*RST:
Full span
Usage:
SCPI confirmed
Manual operation:
See
See
See