
Remote Commands
R&S
®
FSW
736
User Manual 1173.9411.02 ─ 19
Parameters:
<FilterType>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FSW-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FSW-K54) option
P5
5 Pole filters
*RST:
NORM
Refer to the datasheet for available filter bandwidths.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>[:FREQuency]:STARt
<Frequency>
This command defines the start frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the SEM sweep list covers, the R&S
FSW will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the SEM sweep list covers, the R&S
FSW will adjust the start
frequency of the first SEM range and the stop frequency of the last SEM range to
the span
For more information see
chapter 5.6.4.1, "Ranges and Range Settings"
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range>
1...30
Selects the measurement range.
Parameters:
<Frequency>
Numeric value. Note that the minimum frequency range of a
SEM range is 20
Hz.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Manual operation:
See
Configuring and Performing Measurements