Marker Table Display
Defines how the marker information is displayed.
"On"
Displays the marker information in a table in a separate area beneath
the diagram.
"Off"
Displays the marker information within the diagram area.
"Auto"
(Default) Up to two markers are displayed in the diagram area. If
more markers are active, the marker table is displayed automatically.
Remote command:
Marker Stepsize
Defines the size of the steps that the marker position is moved using the rotary knob.
"Standard"
The marker position is moved from pixel to pixel on the display. This
setting is most suitable to move the marker over a larger distance.
"Sweep
Points"
The marker position is moved from one sweep point to the next. This
setting is required for a very precise positioning if more sweep points
are collected than the number of pixels that can be displayed on the
screen. It is the default mode.
Remote command:
Defining a Fixed Reference
Instead of using a reference marker that may vary its position depending on the mea-
surement results, a fixed reference marker can be defined for trace analysis.
When you set the "State" to "On", a vertical and a horizontal red display line are dis-
played, marked as "FXD". The normal marker 1 is activated and set to the peak value
of the trace assigned to marker 1, and a delta marker to the next peak. The fixed refer-
ence marker is set to the position of marker 1 at the peak value. The delta marker
refers to the fixed reference marker.