[SENSe:]POWer:ACHannel:PRESet:RLEVel
.................................................................... 699
.................................................................................................699
CALCulate<n>:MARKer<m>:FUNCtion:POWer:MODE
<Mode>
This command selects the trace display mode for power measurements.
Parameters:
<Mode>
WRITe
The power is calculated from the current trace.
MAXHold
The power is calculated from the current trace and compared
with the previous power value using a maximum algorithm.
Manual operation:
See
CALCulate:MARKer:FUNCtion:POWer<sb>:RESult?
<Measurement>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement; for all other measure-
ments: irrelevant
Configuring and Performing Measurements