Common Analysis and Display Functions
R&S
®
FSW
508
User Manual 1173.9411.02 ─ 19
.............................................................................................. 514
............................................................................................515
Marker Peak List Configuration
.............................................................................517
Deactivating All Marker Functions
.........................................................................519
7.4.3.1
Precise Frequency (Signal Count) Marker
A special marker can be used to determine a particular frequency or time in a mea-
sured signal very accurately. Signal counters are configured in the "Signal Count Con-
fig" dialog box.
To display the "Signal Count Config" dialog box, do one of the following:
●
Press the MKR FUNC key, then select the "Select Marker Function" softkey. Then
select the "Signal Count" button. Select the "Signal Count Config" softkey.
●
In the "Overview", select "Analysis", and switch to the vertical "Marker Function
Config" tab. Then select the "Signal Count" button. Select the "Signal Count Con-
fig" softkey.
For details see
chapter 7.4.1.5, "Performing a Highly Accurate Frequency Measure-
on page 488. A measurement example is described in
ter 5.2.2.2, "Measuring the Signal Frequency Using the Signal Counter"
Signal counters are not available for measurements on I/Q-based data.
.......................................................................................... 508
Signal Count Marker State
Activates or deactivates the special signal count marker function.
When activated, the sweep stops at the reference marker until the signal counter has
delivered a result.
Remote command:
Resolution
Defines the resolution with which the signal is analyzed around the reference marker 1.
Remote command:
Marker Usage