Example:
ROSC:EXT:FREQ 13MHZ
Sets the frequency to 13 MHz.
SOUR:EXT:ROSC:EXT:FREQ 13MHZ
Manual operation:
See
[SENSe:]ROSCillator:OSYNc
<State>
If enabled, a 100 MHz reference signal is provided to the SYNC TRIGGER OUTPUT
connector.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
ROSC:OSYN ON
Manual operation:
See
[SENSe:]ROSCillator:SOURce
<Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S
FSW.
Parameters:
<Source>
INTernal
the internal reference is used (10 MHz)
EXTernal
the external reference from REF INPUT 1..20 MHZ connector is
used with a variable frequency; if none is available, an error flag
is displayed in the status bar
E10
the external reference from REF INPUT 1..20 MHZ connector is
used with a fixed 10 MHZ frequency; if none is available, an
error flag is displayed in the status bar
E100
the external reference from REF INPUT 100 MHZ connector is
used; if none is available, an error flag is displayed in the status
bar
EAUTo
the external reference is used as long as it is available, then the
instrument switches to the internal reference
SYNC
the external reference is used; if none is available, an error flag
is displayed in the status bar
Example:
ROSC:SOUR EXT
Usage:
SCPI confirmed
Configuring the R&S
FSW