Remote Commands
R&S
®
FSW
971
User Manual 1173.9411.02 ─ 19
CALC:MARK:FUNC:FPE:ANN:LAB ON
//--------------- Retrieving results -------------
CALC:MARK:FUNC:FPE:COUN?
CALC:MARK:FUNC:FPE:X?
CALC:MARK:FUNC:FPE:Y?
//------------ Exporting the peak list ------------
MMEM:STOR:PEAK 'PeakList'
Example: Measuring Noise Density
This example demonstrates how to measure noise density using noise markers in a
basic spectrum measurement in a remote environment. It assumes that the basic fre-
quency sweep described in
"Basic Frequency Sweep Measurement for Marker Func-
on page 969 has been performed and thus does not begin by preset-
ting the instrument.
CALC:MARK1:FUNC:NOIS ON
//Switches on noise measurement at marker 1.
INIT;*WAI
//Performs a measurement and waits for it to end
CALC:MARK1:FUNC:NOIS:RES?
//Queries the measured noise level (per Hz bandwidth)
Example: Measuring Phase Noise
This example demonstrates how to measure phase noise using markers in a basic
spectrum measurement in a remote environment. It assumes that the basic frequency
sweep described in
"Basic Frequency Sweep Measurement for Marker Function Exam-
on page 969 has been performed and thus does not begin by presetting the
instrument.
//------------- Configuring the phase noise marker -------------------
DET SAMP
//Switches to Sample detector
CALC:MARK1 ON
//Activates marker1 and sets it to the maximum power level
CALC:DELT:FUNC:PNO ON
//Activates phase noise marker function
CALC:DELT1 ON
CALC:DELT1:X 100kHz
CALC:DELT2 ON
CALC:DELT2:X 500kHz
CALC:DELT3 ON
Analyzing Measurements (Basics)