Common Analysis and Display Functions
R&S
®
FSW
522
User Manual 1173.9411.02 ─ 19
7.4.5 Measurement Example: Measuring Harmonics Using Marker Func-
tions
This measurement example describes how to measure harmonics using the provided
marker functions. Note that this task can be performed much simpler using the Har-
monic Distortion measurement (see
chapter 5.10, "Harmonic Distortion Measurement"
Signal generator settings (e.g. R&S
FSW SMU):
Frequency:
128 MHz
Level:
- 25 dBm
Procedure:
1. Preset the R&S
FSW.
2. Set the center frequency to
128 MHz
.
3. Set the span to
100 kHz
.
The R&S
FSW displays the reference signal with a span of 100 kHz and resolution
bandwidth of 1 kHz.
4. Switch on the marker by pressing the MKR key.
The marker is positioned on the trace maximum.
5. Set the measured signal frequency and the measured level as reference values:
a) Press the MKR FUNC key
b) Press the "Reference Fixed" softkey.
The position of the marker becomes the reference point. The reference point level
is indicated by a horizontal line, the reference point frequency with a vertical line.
At the same time, the delta marker 2 is switched on.