
Remote Commands
R&S
®
FSW
1014
User Manual 1173.9411.02 ─ 19
Usage:
Event
Manual operation:
See
MMEMory:STORe<n>:SGRam
<FileName>
This command exports spectrogram data to an ASCii file.
The file contains the data for every frame in the history buffer. The data corresponding
to a particular frame begins with information about the frame number and the time that
frame was recorded.
Note that, depending on the size of the history buffer, the process of exporting the data
can take a while.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see
chapter 3.1.7, "Protecting Data Using the Secure User Mode"
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:SGR 'Spectrogram'
Copies the spectrogram data to a file.
Manual operation:
See
MMEMory:STORe:SPURious
<FileName>
This command exports the marker peak list available for spurious emission measure-
ments to a file.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see
chapter 3.1.7, "Protecting Data Using the Secure User Mode"
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Managing Settings and Results