Remote Commands
R&S
®
FSW
1025
User Manual 1173.9411.02 ─ 19
Return values:
<Results>
String of data containing the results.
The rows of the self test result table are separated by commas.
Example:
DIAG:SERV:STES:RES?
would return, e.g.
"Total Selftest Status:
PASSED", "Date (dd/mm/yyyy): 09/07/2004 TIME:
16:24:54", "Runtime: 00:06", "...
Usage:
Query only
SOURce:TEMPerature:FRONtend?
This command queries the current frontend temperature of the R&S
FSW.
During self-alignment, the instrument's (frontend) temperature is also measured (as
soon as the instrument has warmed up completely). This temperature is used as a ref-
erence for a continuous temperature check during operation. If the current temperature
deviates from the stored self-alignment temperature by a certain degree, a warning is
displayed in the status bar indicating the resulting deviation in the measured power lev-
els. A status bit in the
STATUs:QUEStionable:TEMPerature
register indicates a
possible deviation.
Return values:
<Temperature>
Temperature in degrees Celsius.
Example:
SOUR:TEMP:FRON?
Queries the temperature of the frontend sensor.
Usage:
Query only
11.10.3 Working with Transducers
The following commands configure and control transducer factors.
Useful commands for transducer managament described elsewhere
●
Remote commands exclusive to transducer management
[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]
........................................... 1026
[SENSe:]CORRection:TRANsducer:COMMent
................................................................1026
[SENSe:]CORRection:TRANsducer:DATA
......................................................................1026
[SENSe:]CORRection:TRANsducer:DELete
....................................................................1026
[SENSe:]CORRection:TRANsducer:SCALing
..................................................................1026
[SENSe:]CORRection:TRANsducer:SELect
.................................................................... 1027
[SENSe:]CORRection:TRANsducer[:STATe]
...................................................................1027
[SENSe:]CORRection:TRANsducer:UNIT
.......................................................................1027
Configuring the R&S
FSW