Remote Commands
R&S
®
FSW
842
User Manual 1173.9411.02 ─ 19
Note that the number of sweep points is limited to 10001 when measuring spurious
emissions.
For EMI measurements, 200001 sweep points are available.
Parameters:
<SweepPoints>
Range:
101 to 100001
*RST:
1001
Example:
SWE:POIN 251
Usage:
SCPI confirmed
Manual operation:
See
[SENSe:]SWEep:TIME
<Time>
This command defines the sweep (or: data capture) time.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths.
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Example:
SWE:TIME 10s
Usage:
SCPI confirmed
Manual operation:
See
See
See
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Usage:
SCPI confirmed
Manual operation:
See
See
See
Setting Basic Measurement Parameters