Common Analysis and Display Functions
R&S
®
FSW
504
User Manual 1173.9411.02 ─ 19
"XY-Search"
Searches in all frames at all positions.
Remote command:
Marker Search Area
Defines which frames the search is performed in.
"Visible"
Only the visible frames are searched.
"Memory"
All frames stored in the memory are searched.
Remote command:
Exclude LO
If activated, restricts the frequency range for the marker search functions.
"ON"
The minimum frequency included in the peak search range is ≥ 5 ×
resolution bandwidth (RBW).
Due to the interference by the first local oscillator to the first inter-
mediate frequency at the input mixer, the LO is represented as a sig-
nal at 0 Hz. To avoid the peak marker jumping to the LO signal at 0
Hz, this frequency is excluded from the peak search.
"OFF"
No restriction to the search range. The frequency 0 Hz is included in
the marker search functions.
Remote command:
Marker Usage