Measurements
R&S
®
FSW
290
User Manual 1173.9411.02 ─ 19
Final test results - Result Summary
The results of the final EMI tests at the active marker frequencies are displayed in the
Result Summary
.
The Result Summary provides the following information:
Label
Description
Type
Marker name
Ref
Reference marker for delta markers
Trace
Assigned trace
X-value
Marker x-value (frequency for final test)
Y-value
Marker y-value (level during inital measurement)
Final Test
Detector used for final EMI test
Line name
Line activated for limit check
Δ
Limit
Delta between measured level and limit line (if active)
The value is colored to indicate the following states:
●
green
: does not exceed limit
●
yellow
: within margin
●
red
: exceeds limit
Final Result
Value measured during final EMI test using specified detector at marker frequency
5.13.3 EMI Measurement Basics
Some background knowledge on basic terms and principles used in EMI measure-
ments is provided here for a better understanding of the required configuration set-
tings.
Resolution Bandwidth and Filter Types
.................................................................290
.....................................................................................291
Frequency Resolution - Sweep Points and Scaling
.............................................. 294
..............................................................................295
.....................................................................................296
Initial Measurement - Peak Search
.......................................................................296
Final Measurement at the Marker Position
........................................................... 297
5.13.3.1
Resolution Bandwidth and Filter Types
EMI testing requires resolution filters with a 6
dB bandwidth. The R&S
FSW EMI mea-
surement adds the following bandwidths that comply to commercial and military stand-
ards to those already available with the base unit:
Commercial (CISPR, FFC etc.)
●
200
Hz
●
9
kHz
Electromagnetic Interference (EMI) Measurement (R&S
FSW-K54)