Remote Commands
R&S
®
FSW
837
User Manual 1173.9411.02 ─ 19
Example:
BAND:AUTO OFF
Switches off the coupling of the resolution bandwidth to the
span.
Usage:
SCPI confirmed
Manual operation:
See
See
[SENSe:]BANDwidth|BWIDth[:RESolution]:RATio
<Ratio>
This command defines the ratio between the resolution bandwidth (Hz) and the span
(Hz).
Note that the ratio defined with the remote command (RBW/span) is reciprocal to that
of the manual operation (span/RBW).
Parameters:
<Ratio>
Range:
0.0001 to 1
*RST:
0.01
Example:
BAND:RAT 0.1
Usage:
SCPI confirmed
Manual operation:
See
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE
<FilterType>
This command selects the resolution filter type.
When you change the filter type, the command selects the next larger filter bandwidth if
the same bandwidth is unavailable for that filter.
The EMI-specific filter types are available if the EMI (R&S
FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see
ter 5.13.3.1, "Resolution Bandwidth and Filter Types"
Setting Basic Measurement Parameters