Low plasma power. Call ASP Technical
Service.
Problem/Condition
RF delivered power is less than 365 W but greater than 200 W.
Action
F
Remove the shelves and run an empty chamber cycle. If completed
correctly, reinstall the shelves.
F
Check the RF subsystem for malfunction.
Moisture in load. Dry and repackage
load. Restart sterilizer.
Problem/Condition
Pressure rises of 50 mTorr or more during pump down in the vacuum
phase in the pressure range of approximately 4 Torr to 1 Torr are
indicative of water freezing.
Action
F
Handle the load with gloves as H2O2 may be present.
F
Dry load.
F
Repackage the load.
F
Restart the sterilizer.
NO CASSETTE PRESS=XXX.XX MIN
There is no printed message for this condition.
Problem/Condition
No cassette was loaded at the start of the cycle or an unacceptable
cassette was inserted during the cycle.
Action
Insert a new cassette.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-15