Replace the Electrode and Spacers
1. Place the electrode into the chamber. Starting with the ones on the
bottom, insert 18 new electrode spacers and new shelf stops.
Alternate the tabs up and down.
2. Insert the lower shelf supports into the chamber and slide them
towards the back of the system. Replace the two sets of two screws
(two each located at the front of each lower shelf support) fastening
the lower shelf supports to the electrode assembly.
3. Insert the upper shelf supports into the chamber and slide them
towards the back of the system. Replace the two sets of two screws
(one located at the back and one located at the middle of each upper
shelf support) fastening the upper shelf supports to the electrode
assembly.
4. Slide the upper left and right joiners towards the back of the system.
5. Replace the Allen screw (located at the left, lower middle of the
electrode) fastening the electrode to the chamber feedthrough.
6. Insert the upper and lower shelves into the chamber. (The diagram
shows the shelves, shelf supports and spacers correctly installed.)
Figure 7-6. Electrode Spacers/Shelf Support Locations
Door and Chamber Subsystem
Removal/Replacement
7-14
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......