RF Calibration
IMPORTANT!:
Your first step when troubleshooting a system is to verify
that the system is
properly connected to the power source
, and to
make sure all the
subsystems are properly connected to the AC
system and all circuit breakers are on and functioning normally. Once
you have verified these things, proceed with your troubleshooting steps.
RF Calibration is performed during the AutoTest procedures. See the RF
Test in the AutoTest Appendix.
RF Removal/Replacement
The RF enclosure, the RF Match and the feedthrough are field
replaceable unit. The RF generator is not replaceable. If the generator
needs to be replaced, the whole enclosure must be replaced. The
feedthrough replacement is found in Chapter 7.
Remove the RF Enclosure
1. Unplug the RF generator from the controller.
2. Unplug the unit from the AC power.
3. Unscrew the RF enclosure from the frame and remove it from the
system. See the enclosure diagram
To replace the RF enclosure do the following:
1. Attach the RF enclosure to the frame using the 6 screws. See the
diagram.
2. Plug in the RF generator into the controller.
3. Plug in the AC power.
Remove the RF Match
1. Remove the 5 screws securing the RF match cover. Remove the RF
match cover.
2. Remove the nut and screw securing the RF match to the feedthrough.
See the RF diagram.
To replace the RF match, do the following:
Radio Frequency (RF) Subsystem
RF Calibration
10-4
STERRAD® 100 Sterilizer
ASP Service Guide rev. D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......