If the cycle does not run successfully, perform the tests below.
F
Perform the vacuum/RF test.
F
Perform the Baratron zero-shift test.
F
Perform leak back test.
F
Replace vacuum pump
Vacuum not low enough for injection.
Repackage load. Restart sterilizer.
Problem/Condition
F
Vacuum did not reach 1 Torr within 9 minutes.
F
Vacuum did not reach 300 mTorr after an auto-restart.
Action
F
Check for moisture in the load.
F
Remove any cloth, paper or foam items from the load.
F
Repackage the load.
F
Restart the sterilizer.
Vacuum System Interrupted Call ASP
Technical Service
Problem/Condition
The chamber is still at atmospheric pressure after the first 30 seconds of
vacuum.
Action
F
Test the vacuum system
F
Check power for proper CBA phase rotation.
F
Test the vacuum switch
F
Perform vacuum/RF test.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-39