Block 2 Injector Valve/Vaporizer
Removal
4
4
Note:
The injector pump must be removed, or loosened, prior to
removing the injector valve.
1. Remove the 2 screws attaching the valve to the vaporizer. Lift the
valve up.
2. Disconnect the vaporizer heater cable.
3. Insert the spanner wrench into the retaining ring holes, turn it
counter-clockwise.
4. Place your hand inside the chamber to catch the vaporizer. Press
down on the vaporizer to drop it into the chamber.
Replace the Injector Valve
1. Make sure the injector boss is clean so as not to damage the new
O-ring. Insert the injector valve through the opening from inside the
chamber until the O-ring catches and holds the valve in place.
2. Reinsert the screws and tighten with the hex wrench. Than back-off
3 complete turns. Seat the valve by pushing down.
3. This step is for Block 2 systems: with the heater assembly in place
insert the vaporizer through the top opening from inside the chamber.
4. Place the spacers in their proper grooves (one is large than the other).
Insert the O-ring and hand tighten.
5. Once the vaporizer is hand tight, use the spanner wrench to tighten
completely.
6. Reinstall the injector valve into the vaporizer and reinsert the screws;
do not tighten at this point.
Injection Subsystem
Injection Subsystem Removal/Replacement
5-12
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......