4
Note:
Lower the injector valve to obtain required clearance to
insert tool over injector valve.
2. Raise the injector valve by turning the injector valve adjustment
screws clockwise until no gap exists between the clamp seal block
and the cassette gap tool (the O-ring should fit into the cutaway of
the tool). Adjust the valve height to a point where the cassette gap
tool is difficult to remove.
3. Remove the cassette gap tool.
Injector Valve Height-Block 2
1. Fully extend the injector pump cam and leave it extended. From the
back of the platen, insert the cassette gap tool (injector valve gage).
4
4
Note:
Using an Allen wrench, raise the injector pump clamp
seal block by rotating the Allen screw on the knob at the
bottom of the seal base. This creates the required
clearance needed to insert the tool over injector valve.
2. Loosen the set screw on the injector seal block adjustment knob.
3. Push platen down to allow the tool to advance on top of the injector
valve.
4. Verify that the seal block O-ring and needle seals clear the opening in
the cassette gap tool.
5. Lower the seal until no gap exists between the seal block and the tool
or between the needle seal retainer plate and the tool. Lock the seal
with the set screw.
6. Remove the cassette gap tool.
Injector Pump Valve Adjustment
This calibration/adjustment is only applicable to
Block 2
devices.
1. Loosen
but do not remove
the two rear screws mounting the injector
pump to the injector boss.
2. Adjust the locknuts and mounting hardware at the front of the
injector pump so that an inserted or ejected cassette passes evenly
centered through the front panel cassette door without sticking or
binding on any surface. Secure the locknuts and mounting screws.
3. Using the cassette gap tool, re-check the injector valve setting from
injector pump clamp. Perform the check with the cam fully extended
and adjust if necessary so that tool is fully contacting seal clamp
plate of injector valve and clamp base (not the O-ring) of the injector
pump clamp.
Injection Subsystem Calibration
Injection Subsystem
ASP Service Guide rev D
STERRAD® 100 Sterilizer
5-5
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......