Introduction
This section is about the AutoTest board. The AutoTest Board enables
service personnel to perform calibration and product certification
procedures more easily. The procedures featured are those procedures
that are benefited by using the AutoTest board. Some procedures are
the same as those used with the Service Test board, others automate
manual service test procedures.
AutoTest is menu driven and selections are made from the control panel
display LCD by using the function keys on the control panel. The use of
the function keys as displayed on the LCD is as follows:
F
START
-Press
START
.
F
CANCEL
-Press
CANCEL
.
F
LONG CANCEL
-Press and hold
CANCEL
for 3 seconds.
F
OPEN-
Press
OPEN DOOR.
F
CLOSE-
Press
CLOSE DOOR
.
F
LONG
CLOSE-
Press and hold
CLOSE DOOR
for 3 seconds.
Overview
AutoTest has a long series of selectable tests. If a test shows an anomaly,
you should determine the source of the problem, fix it according to the
corrective actions and then restart the test. Some of the tests contained
in the firmware are not performed. They are presented for reference
only. Differences in configurations are shown where applicable. The
tests are presented as follows:
Test Name
This heading indicates what subsystem is being tested.
Display
Shows the messages displayed on the LCD read-out.
SAMPLE LCD Display
AutoTest
Overview
B-2
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......