Vacuum Insufficient. See Op Manual
Appendix A
Problem/Condition
The pressure did not reach the value necessary for pre-plasma within 12
minutes. Load may be out-gassing.
Action
F
Run an empty chamber cycle. If it completes successfully, the
problem is in the load.
F
Repackage the load and restart the sterilizer.
F
If the empty chamber cycle is unsuccessful, perform a leak back test.
F
Check vacuum pump and add oil if needed.
F
Check PM log and perform PM if needed.
F
Replace throttle valve.
F
Replace vacuum pump.
Vacuum Insufficient for Plasma. Call ASP
Technical Service.
Problem/Condition
For the second consecutive cycle, the vacuum did not reach 525 mTorr
in the first 3 minutes of the plasma stage. Load may be out-gassing.
Action
F
Run an empty chamber cycle. If it completes successfully, the
problem is in the load.
F
If the empty chamber cycle is unsuccessful, perform a leak back test.
F
Check the vacuum pump and add oil if needed
F
Check the PM log and perform PM if needed
F
Replace the throttle valve
F
Replace the vacuum pump
F
Replace vent valve.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-37
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......