High/Low Pressure Range Divider Voltage
CAUTION: Be sure to use the correct leads on Block 2 systems to
avoid shorting out the power supply.
1. Open the control enclosure and interrupt power to the enclosure
using the safety interlock. Remove the microprocessor board, and
insert the AutoTest board.
2. Remove the connector from the Baratron pressure transducer, and
attach the voltage divider adjustment tool to the harness connector.
3. Apply power to the enclosure using the safety interlock.
4. Perform the Analog Test.
5. At the completion of the test, interrupt power using the safety
interlock, remove the AutoTest board and install the system
microprocessor board. Remove the adjustment tool and reconnect
the cable to the pressure transducer. Apply power to enclosure using
the safety interlock. Close the enclosure.
User Subsystem Replacement
Procedures
Remove/Replace the Printer
1. Unplug the ribbon cable. Remove the paper roll.
2. Unscrew the two screws holding the printer to the frame.
3. Lift up on the printer to remove it.
4. Replace the printer by tipping it into place and replacing the screws.
5. Plug in the ribbon cable. Replace the paper roll.
User Interface and Control Subsystems
User Subsystem Replacement Procedures
6-22
STERRAD® 100 Sterilizer
ASP Service Guide rev D.
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......