F
Replace the RF match.
F
Replace the RF enclosure.
F
Check the microprocessor board and replace if necessary.
4
Note:
The RF match and enclosure may need to be replaced at
the same time.
Leak Identification Tips
Do any one of the following to identify the source of the leak.
1. Take the sight glass apart (not present on Block 2) and inspect it for
damage.
2. Check the KF-40, KF-16 and KF-25 O-rings and clamps on the
vacuum subsystem and chamber. If they are cracked or loose,
replace them.
3. Check the bellows for cracks, replace if damaged.
4. Check the door seal O-ring and replace if necessary.
5. Check the vent valve by isolating it with the plastic cap that comes
with the throttle valve.
6. Check the vacuum switch and the O-ring that seals it. It may be too
tight or too loose.
7. Isolate the throttle valve by stretching a bellows and installing it
between the intake filter and the chamber connector. Replace the
throttle valve if necessary.
8. Check the RF feedthrough and replace if necessary.
9. Isolate the vacuum pump by manually closing the throttle valve
during pump down. Electrically disconnect the throttle valve. If the
pressure does not rise when the throttle valve is closed, the leak is
below the throttle valve—either in the pump intake filter or the pump
itself. If the leak, continues the vent valve may be the problem..
System Messages
Using System Messages
3-46
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......