Temperature over 60 degrees. Call ASP
Technical Service.
Problem/Condition
The temperature of the door or chamber has risen above 60° C.
Action
F
Replace the
s
olid state relay board
F
Replace the analog board
F
Replace the thermistor.
Temperature over threshold. Call ASP
Technical Service
Problem/Condition
The temperature of the door or chamber has risen above 60° C.
Action
F
Replace solid state relay board
F
Replace analog board
F
Replace thermistor.
TEMP. OVER THRESHOLD. NO MORE
CYCLES CAN RUN
Temperature Over Threshold. Call ASP Technical Service.
Problem/Condition
The temperature of the door or chamber has risen above the specified
limits.
Action
F
Replace state relay board.
F
Replace analog board.
F
Replace thermistor.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-35