No RF Power in Vacuum. Call ASP
Technical Service
Problem/Condition
RF delivered power is less than 100 W.
Action
F
Check the load and make sure it is not in contact with the electrode.
F
Check the shelves and make sure they are not in contact with the
door, or the chamber.
F
Check the RF power and replace if necessary.
F
Replace the RF amplifier
F
Replace the RF match.
F
Replace the electrode
F
Replace the microprocessor board.
No RF Power. Call ASP Technical Service
Problem/Condition
RF delivered power is less than 100 W.
Action
F
Check the load and make sure it is not in contact with the electrode.
F
Check the shelves and make sure they are not in contact with the
door, or the chamber.
F
Check the RF power and replace if necessary.
F
Replace the RF amplifier
F
Replace the RF match.
F
Replace the electrode
F
Replace the microprocessor board.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-17